Live Results Eijsden
To overview 2
EasyClus check & Info of last file

----- Processed at : 24-Apr-2025 06:27:02 -----

1. The cytometric analysis is performed on machine Lobith
2. The CytoBuoy instrument serial-nr is CS-2021-106
3. The CytoUSB software used is CytoUSB Version: 5.11.3.1
4. LASER : OBIS LS 488-60 : Used at 50 mW
5. LASER : OBIS LS 552-60 : Used at 50 mW
6. The filename is EIJ2_SWS100_FLO100_2_uls_beads_600s 2025-04-24 06h12.cyz
7. The analysis is performed on 24-04-2025 time 06:16:07
8. The number of measured particles is 492
9. The total concentration found (e.g. only SWS trigg.) is 117 particles/microliter
10. The concentration found (all used triggers) is 1 particles/microliter
11. The preconcentration (before analysis) found (all used triggers) is 1 particles/microliter
12. The flush time found (between samples) is 130 seconds
13. The effective analysis volume is 849.8538 microliter
14. The pumped volume is 852 microliter
15. The particle-rate is 167 particles passing the laser beam/second
16. The used sample flow rate is 1.4245 microliter/second
17. Trigger 1: SWS HS at threshold level 100 mV
18. Trigger 2: Maximum Fl Orange at threshold level 100
19. Temperature sensor PMT is : 25.2 C
20. Temperature sensor Sheath is : 20.3 C
21. Temperature sensor Laser is : 24.6 C
22. Temperature sensor System is : 22.7 C
23. Pressure sensor difference is : -448 mBar
24. Pressure sensor absolute is : 342 mBar
25. Sample pump 1: Last calibration of the sample pump was : 05-Dec-2024
26. Alignment check 1 by curvature detector is 1.088 (1=perfect)
27. Alignment check 2 by curvature detector is 70.1% (>70%=perfect)
28. Saturated signals (flattened peaks) occur when the dynamic range of the (HS)detector is reached
29. FWS: Nr of saturated signals is 7 (0% of 492 measured particles)
30. FWS: Maximum signal height found is 14464 mV
31. FWS: Nr of particles with signal height <10 mV is high (0 part. (0 %))
32. SWS HS: Nr of saturated signals is 1 (0% of 492 measured particles)
33. SWS HS: Maximum signal height found is 7232 mV
34. SWS HS: Nr of particles with signal height <10 mV is high (0 part. (0 %))
35. FL Yellow HS: Nr of saturated signals is 1 (0% of 492 measured particles)
36. FL Yellow HS: Maximum signal height found is 7360 mV
37. FL Yellow HS: Nr of particles with signal height <10 mV is high (0 part. (0 %))
38. FL Orange HS: Nr of saturated signals is 3 (0% of 492 measured particles)
39. FL Orange HS: Maximum signal height found is 7360 mV
40. FL Orange HS: Nr of particles with signal height <10 mV is high (0 part. (0 %))
41. FL Red HS: Nr of saturated signals is 6 (0% of 492 measured particles)
42. FL Red HS: Maximum signal height found is 7360 mV
43. FL Red HS: Nr of particles with signal height <10 mV is high (0 part. (0 %))
44. FWS: Nr of block shaped signals is 27 (5.4878%)
45. FL Yellow HS: Nr of block shaped signals is 9 (1.8293%)
46. FL Orange HS: Nr of block shaped signals is 9 (1.8293%)
47. FL Red HS: Nr of block shaped signals is 9 (1.8293%)
48. FL Red LS: Nr of block shaped signals is 7 (1.4228%)
49. SWS HS: Nr of block shaped signals is 5 (1.0163%)
50. Suspect nr: 0=low, 10=high
51. Maximum FWS: suspect = 4
52. Maximum SWS HS: suspect = 4
53. Maximum SWS LS: suspect = 0
54. Maximum FL Yellow HS: suspect = 3
55. Maximum FL Yellow LS: suspect = 0
56. Maximum FL Orange HS: suspect = 5
57. Maximum FL Red HS: suspect = 3
58. Maximum FL Red LS: suspect = 0
59. SWS HS: PMT level is 49
60. SWS LS: PMT level is 48
61. FL Yellow HS: PMT level is 94
62. FL Yellow LS: PMT level is 76
63. Fl Orange: PMT level is 105
64. FL Red HS: PMT level is 110
65. FL Red LS: PMT level is 81
66. IIF: Imaging in Flow pictures available. Nr of images is 250

-------------------------- Warnings info --------------------------
1.01 Warning: The number of saved particles in this analysis is quite low and
1.01 Warning: you might miss (statistically) bigger particles in your sample
1.01.1 Solution - Increase number of measured particles. Increase setted time OR after setted number of analyzed particles (in CytoUSB)
1.11 Warning: This threshold level is quite high
1.11 Warning: Take care you don't miss smaller particles, because they will not be triggered at this level
1.11.1 Solution - Set lower threshold trigger level
1.11.2 Solution - If you need to put the level that high to measure something, it might be causeed by increased noise.
1.11.3 Solution - Increased noise is usually caused by dirty cuvette (inner) walls/tubing/sheath fluid - So, clean the cuvette carefully, change tubing, use new sheath
1.21 Warning: Last calibration of sample pump towards analysis date was more than 90 days. (05-Dec-2024)
1.21.1 Solution - Check if the pumped analysis volume is the real volume by weighing
1.24 Warning: SWS LS: Maximum is 340 mV if HS is around 4500 mV and too high.
1.24.1 Solution - Put LS PMT lower. Maximum LS around 50 mV (if HS is around 4500 mV).
1.24 Warning: SWS LS: HS-LS ratio is 67 and too low.
1.24.1 Solution - Put LS PMT lower. Ratio HS/LS should be around 100.
1.24 Warning: FL Yellow LS: HS-LS ratio is 55 and too low.
1.24.1 Solution - Put LS PMT lower. Ratio HS/LS should be around 100.
1.24 Warning: FL Yellow LS: HS-LS ratio is 46 and too low.
1.24.1 Solution - Put LS PMT lower. Ratio HS/LS should be around 100.
1.24 Warning: FL Red LS: HS-LS ratio is 68 and too low.
1.24.1 Solution - Put LS PMT lower. Ratio HS/LS should be around 100.
1.28 Warning: FWS: A relative large amount of particles are (rather) block shaped (5%)
1.28.1 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.28.2 Solution - Decrease, set lower FWS PMT voltage in CytoUSB (make less sensitive)
1.25 Warning: Suspected particles data blocks found!
1.29.1: Maximum FWS: suspect = 4
1.29.11 Solution - Particle rate too high?
1.29.21 Solution - Trigger too sensitive?
1.29.31 Solution - Details look at 'DATA DIAGNOSIS' in 'Check CytoSense-Diagnosis-Meta Data'
1.29.2: Maximum SWS HS: suspect = 4
1.29.12 Solution - Particle rate too high?
1.29.22 Solution - Trigger too sensitive?
1.29.32 Solution - Details look at 'DATA DIAGNOSIS' in 'Check CytoSense-Diagnosis-Meta Data'
1.29.3: Maximum FL Yellow HS: suspect = 3
1.29.13 Solution - Particle rate too high?
1.29.23 Solution - Trigger too sensitive?
1.29.33 Solution - Details look at 'DATA DIAGNOSIS' in 'Check CytoSense-Diagnosis-Meta Data'
1.29.4: Maximum FL Orange HS: suspect = 5
1.29.14 Solution - Particle rate too high?
1.29.24 Solution - Trigger too sensitive?
1.29.34 Solution - Details look at 'DATA DIAGNOSIS' in 'Check CytoSense-Diagnosis-Meta Data'
1.29.5: Maximum FL Red HS: suspect = 3
1.29.15 Solution - Particle rate too high?
1.29.25 Solution - Trigger too sensitive?
1.29.35 Solution - Details look at 'DATA DIAGNOSIS' in 'Check CytoSense-Diagnosis-Meta Data'
Disclaimer
Any use of the provided data and results or information provided by this website is on full own risk and responsibility. Those who gave access to this website or those who provided the results represented on the website cannot be held responsible for any damage or consequences that could arise from the use of this information. This LIVE system is empowered by Rijkswaterstaat Laboratorium, CytoBuoy BV and Thomas Rutten Projects. TRP © 2024